3.3.5 Alarm System

Repeats Until Error Message
ID Error Message
Actions after Error
LLC Test
Code
Code
1.
Prerequisites for Test:
No alarm activation
2.
Test code = 16 to LLS
3.
1610x
OSD LED OFF, Sound OFF
No alarm activation
Timeout 60 s
4.
5.
1620x
Wait for D24OFF and B24OFF (timeout 60 s since start
of code 100)
6.
7.
1630x
Wait for D24ON and B24ON (timeout 10 s)
8.
1630x
D24ON and B24ON:
Alarm activation LLC/TLC
Activation test alarm with sound and OSD LED red (ID
1002)
9.
10.
1640x
Wait for D24OFF (timeout 60 s since start of code 200)
11.
12.
1650x
Wait for D24ON (timeout 10 s)
13.
1699
Completed
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
3 . R e p a i r I n s t r u c t i o n s
16
Test Code
1
1167: Sound + LED test failed
None (test is repeated immediately)
LLC Test Steps
LLC
LLS Test Steps
LLS
Test code = 16 back to LLC
Start condition for the test:
1.
„OSD-LED OFF" detected
2.
„Sound OFF" detected
Activation of the red OSD LED
Wait for feedback of the activation and switch-off of
the LED
Wait for LED off
Acknowledge with switch-off of the power supply
trigger, i.e. switch-off of VGD and VGB
Trigger ON
Condition for the test:
1.
„OSD LED red left and right ON" detected
2.
„Sound ON" detected
Acknowledge with switch-off VGD
VGD ON
1 / 2 0 1 0
3 - 11
LLS Test
Code
Code
16001
16002
16003
16004
16005 –
16010
16011
16021
16022 –
16027
16028