Calibration Degassing Temperature Sensor Tse/Heater Inlet Temperature Sensor Tshe - B. Braun Dialog+ Service Manual

Sw 9.xx
Table of Contents

Advertisement

+
Dialog
SW 9.xx
4.8.3.7
Calibration Degassing Temperature Sensor TSE/
Heater Inlet Temperature Sensor TSHE
Prior to Calibration
DM V
VVBE
H
TSE
EK
VB W T
W AB
T
PE
TSH
A
D
T
TSHE
FEP
VEB
EP
8
Fig.: Temperature Circulation
KP/BICP
VEB
EP
Fig.: Original Tubing Water Sub-Rack for TSE/TSHE
BA-TE-DE08W M.KAY
Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
4 . T S M S e r v i c e P r o g r a m
• Position the heater inlet temperature sensor TSHE next to TSE (see figure
redirect heater inlet temperature sensor TSHE, next page)
• Use calibrated measurement instrument
• Wait until Dialog has reached working temperature
The degassing temperature sensor TSE and the heater inlet temperature sensor
TSHE are calibrated together. The following calibration points are used:
TSE / TSHE: 30 °C, 40 °C, 85 °C
VLA
RVDA
SKBS
DBK*
T
TSBIC
VBICP*
VBKO*
VVB*
VBKS
RVBO*
. ..
NSVB
WT
H
DBK
TSE
WAB
TSHE
RVDA
FEP
DMV
VVBE
H
EK
VB WT
WB
T
PE
TSH
A
D
T
TSHE
FEP
VEB
EP
8
Fig.: Loop-In Point and TSHE in Line with TSE for Calibration
KP/BICP
VEB
EP
Fig.: Water Sub-Rack with TSHE in Line with TSE for Calibration
1 / 2 0 1 0
Measurement
Instrument
DBK*
T
T
TSHE
TSE
VBKO*
VVB*
VBKS
RVBO*
Measurement
Instrument
NSVB
WT
H
DBK
TSHE
TSE
WAB
RVDA
FEP
B. Braun Avitum AG
4 - 128
ml/min
o
C
mS/cm
ml/min
o
C
mS/cm
Measurement
Sensor
Keep tubings as short
as possible