Handler Interface Output Test; Handler Interface Output Order - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

10. Set the measurement parameter to current (I).
a. Press
.
b. Select using
11. Press the IN5 switch on the Handler Interface Tester, and con rm that the 4349B is
triggered.

Handler Interface Output Test.

12. Start the handler interface output test with the following procedure.
a. Press
b. Select
using
c. Select
using
13. Set the LED PW switch on the Handler Interface Tester to ON .
Caution
Don't set the LED PW switch to ON , except when doing the handler interface
output test. The LEDs may cause the 4349B to shut-down during normal
operation.
14. Con rm that the LEDs on the Handler Interface Tester light in the order shown in
Figure 9-11, in accordance with the 4349B display.
15. Set the LED PW switch on the Handler Interface Tester to OFF.
16. Press
twice to exit from the test mode.
9-24
Maintenance
or
and press
to display the con guration setting menu.
or
and press
or
and press
Figure 9-11. Handler interface Output Order
.
.
.