Reading the Limit Data of Contact Check Function

1. Select the desired channel.
2. Press
.
3. Select
using
4. Select
using
5. The limit value is displayed.
The limit data ( Limit ) is calculated by the following equation.
Where:
Cs
Stray capacitance at open measurement (Value returned from the GPIB
command
O set
Data entered by
is set to `0''(default) in normal measurement. In case of a parallel capacitor
measurement,
Note
Set the o set value to more than 0.
or
, then press
or
, then press
3
Limit = C
1 : 035 + 0 : 40
s
will be set to the capacitance of the parallel capacitor.
.
.
3
10
+ Off set
12
)
Operating the 4349B
2-9