Resistance Measurement Accuracy Test Settings - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B
8. Press the 3458A Multimeter's
voltage calibrator is used, skip this step.
9. Set the RC Box resistor to 10
10. Set the DC power supply (Voltage Calibrator) to 100 V.
11. Enter the multimeter reading (voltage calibrator setting) into the 4349B using the following
procedure.
a. Press
to display the voltage setting menu.
b. Enter the desired voltage using the numeric keys, and press
c. Con rm that the desired voltage is displayed.
12. Press
to start measurement.
13. Record the 4349B reading on the calculation sheet.
14. Set the DC power supply (voltage calibrator) to 0 V.
15. Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
16. Perform this test for all settings listed in Table 9-4. The DC power supply (voltage
calibrator) must be set to 0 V after each test. In the test for 10
triaxial cable and BNC cable to the 10
connector position.
Table 9-4. Resistance Measurement Accuracy Test Settings
17. Press
to select channel 2.
18. Test channel 2 using the steps 4 to 16.
19. Test channel 3 and channel 4 using the same procedure. If the 4349B is equipped with
Option 001 (Two Test Channels), skip this step.
to set the measurement mode to DC voltage. If the
4
DCV
5
.
6
terminals and set the 10
11
Resistor
Voltage
Measurement
Setting
Setting
10
100 V
6
10
100 V
7
10
100 V
8
10
100 V
8
10
100 V
9
10
100 V
10
10
100 V
11
10
100 V
11
10
10 V
11
.
resistor, connect the
11
switch to the BNC
11
Time
10 ms
30 ms
30 ms
400 ms
30 ms
30 ms
30 ms
10 ms
30 ms
Maintenance
9-11