Resistance Measurement Accuracy Test; Speci Cation; Test Equipment; Procedure - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B

Resistance Measurement Accuracy Test

The 4349B measures calibrated RC Box with an external DC voltage source, and its readings are
compared with the RC Box's calibration values.
Speci cation
Basic Measurement Accuracy:

Test Equipment

Description
Multimeter
System DC Power Supply
RC Box
Dual Banana-BNC(m) Cable
The DC voltage calibrator listed in Table 9-2 can be used as a substitute.
1

Procedure

1. Record the 16340A RC Box calibration values into the calculation sheet.
2. Reset the 4349B using the following procedure.
a. Press
b. Select
using
3. Press
three times to set the measurement time to 10 ms. The measurement time
setting is indicated by the Meas Time annunciator(
4. Press
to set the trigger mode to Manual. The trigger mode is indicated by the
Trigger annunciator(
5. Set up the equipment as shown in Figure 9-7.
Figure 9-7. Resistance Measurement Accuracy Test Setup
6
2 % (See Chapter 8 General Information for details.)
Recommended Model
3458A
1
1
6634A
16340A
PN 11001-60001
to display the system reset settings.
or
and press
).
9
.
).
9
Maintenance
9-9