Contact Check Functional Test; Test Equipment; Procedure; Contact Check Functional Test Setup - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B

Contact Check Functional Test

The 4349B's contact check function is con rmed.

Test Equipment

Description
SMU Test Adapter
BNC(m)-BNC(m) Adapter

Procedure

1. Reset the 4349B with the following procedure.
a. Press
b. Select
using
2. Set up the equipment as shown in Figure 9-12.
3. Set the SMU Test Adapter switch to Vo position.
4. Set the measurement range to 10  A with the following procedure.
a. Press
to display the measurement range setup menu.
b. Select
 using
5. Set the measurement parameter to current (I).
a. Press
.
b. Select using
6. Press
Recommended Model
PN 04145-65001
PN 1250-0216
to display the system reset settings.
or
and press
Figure 9-12. Contact Check Functional Test Setup
or
and press
or
and press
to perform the OPEN correction.
.
.
.
Maintenance
9-25