Measurable Capacitance Range
Measurable Current Range and Measurable Capacitance Range
Range
100 pA
1 nA
10 nA
100 nA
1 A
10 A
100 A
Current Measurement Range shows the range where the 4349B's specied accuracy is
applied. This range is applicable for a pure resistor (non-capacitive device) measurement.
Measurable Capacitance Range shows the capacitance range that can be measured in each
of the 4349B's measurement ranges.
8-8
General Information
Measurement
Current measurement
Time
10 ms
30 ms
1 pA to 145 pA
100 ms
1.4 pA to 145 pA
400 ms
1.4 pA to 145 pA
10 ms
10 pA
30 ms
2 pA to 1.45 nA
100 ms
5 pA to 1.45 nA
400 ms
5 pA to 1.45 nA
10 ms
30 pA to 14.5 nA
30 ms
11 pA to 14.5 nA
100 ms
41 pA to 14.5 nA
400 ms
41 pA to 14.5 nA
10 ms
200 pA to 145 nA
30 ms
100 pA to 145 nA
100 ms
400 pA to 145 nA
400 ms
400 pA to 145 nA
10 ms
2 nA to 1.45 A
30 ms
1 nA to 1.45 A
100 ms
4 nA to 1.45 nA
400 ms
4 nA to 1.45 nA
10 ms
20 nA to 14.5 A
30 ms
10 nA to 14.5 A
100 ms
400 ms
10 ms
120 nA to 100 A
30 ms
100 ms
400 ms
Table 8-2.
range
|
1.45 nA
|
|
|
|
|
Measurable Capacitance
Range
|
1 nF
10 nF
10 nF
1 nF
10 nF
100 nF
100 nF
10 nF
100 nF
1 F
1 F
100 nF
1 F
10 F
10 F
1 F
10 F
100 F
100 F
10 F
100 F
|
|
100 F
|
|
|
4349B