Agilent Technologies 4349B Manual page 121

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B
FETCh? Query
:FETCh?
The
query retrieves the measurements taken by the
them into the 4349B's output bu er. (Query only)
Query response is,
< stat1 > , < data1 > , < comp1 > , < stat2 > , < data2 > , < comp2 > , < stat3 > , < data3 > , < comp3 > ,
< stat4 > , < data4 > , < comp4 >
Where,
< stat1-4 > measurement status of each channel
0 : normal
1 : overload
2 : no-contact
< data1-4 > measurement data of each channel
< comp1-4 > comparison result of each channel (no output when the comparator
function is OFF)
1 : In
2 : High
4 : Low
8 : No-Contact
Format is speci ed by
For more information, refer \To Trigger a Measurement" in Chapter 4.