Change 2
Change Table C-1 and Table C-2 to Table A-3 and Table A-4 respectively.
Table A-3. OVLD/N.C. Operations (Meas. parameter: R)
Display
Data
Mode
OVLD
OVLD
(Overload)
N.C.
N.C.
(No-Contact)
OVLD
&
N.C.
N.C.
Refer to :FETCh? in Chapter 5
1
Measurement status (0:Normal, 1:Overload, 2:No-Contact)
2
Comparison result (1:In, 2:High, 4:Low, 8:No-Contact)
3
Table A-4. OVLD/N.C. Operations(Meas. parameter: I)
Display
Data
Mode
OVLD
OVLD
(Overload)
N.C.
N.C.
(No-Contact)
OVLD
&
N.C.
N.C.
Refer to :FETCh? in Chapter 5
1
Measurement status (0:Normal, 1:Overload, 2:No-Contact)
2
Comparison result (1:In, 2:High, 4:Low, 8:No-Contact)
3
A-4
Manual Changes
Handler Output
Comprtr
Mode
LOW
/LO1-4
N.C.
/NO CONTACT1-4
N.C.
/NO CONTACT1-4
Handler Output
Comprtr
Mode
HIGH
/HI1-4
N.C.
/NO CONTACT1-4
N.C.
/NO CONTACT1-4
GPIB Output
1
<
>
2
stat1-4
: 1
<
>
data1-4
: 9.9E37 Select an appropriate measurement range.
<
>
3
comp1-4
: 4
<
>
2
stat1-4
: 2
Cancel the bad contact between DUT and
<
>
data1-4
: 9.9E37 contact-pin.
<
>
3
comp1-4
: 8
<
>
2
stat1-4
: 3
Select an appropriate measurement range.
<
>
data1-4
: 9.9E37 Cancel the bad contact between DUT and
<
>
3
comp1-4
: 8
contact-pin.
GPIB Output
1
<
>
2
stat1-4
: 1
<
>
data1-4
: 9.9E37 Select an appropriate measurement range.
<
>
3
comp1-4
: 2
<
>
2
stat1-4
: 2
Cancel the bad contact between DUT and
<
>
data1-4
: 9.9E37 contact-pin.
<
>
3
comp1-4
: 8
<
>
2
stat1-4
: 3
Select an appropriate measurement range.
<
>
data1-4
: 9.9E37 Cancel the bad contact between DUT and
<
>
3
comp1-4
: 8
contact-pin.