Initial Inspection

Warning
To avoid hazardous electrical shock, do not turn on the 4349B when there
are signs of shipping damage to any portion of the outer enclosure (for
example, covers, panel, or display)
Inspect the shipping container for damage. If the shipping container or cushioning material
is damaged, it should be kept until the contents of the shipment have been checked for
completeness and the 4349B has been checked mechanically and electrically. The contents of
the shipment should be as follows:
4349B 4-Channel High Resistance Meter
Power cable
Operation Manual (Option ABA, This book)
If the contents are incomplete, if there is mechanical damage or defect, or if the analyzer does
not pass the power-on selftests, notify the nearest Agilent Technologies oce. If the shipping
container is damaged, or the cushioning material shows signs of unusual stress, notify the
carrier as well as the Agilent Technologies oce. Keep the shipping materials for the carrier's
inspection.

Providing clearance to dissipate heat at installation site

To ensure the speci cations and measurement accuracy of the product, you must keep ambient
temperature around the product within the speci ed range by providing appropriate cooling
clearance around the product or, for the rackmount type, by forcefully air-cooling inside
the rack housing. For information on ambient temperature to satisfy the speci cations and
measurement accuracy of the product, refer to Chapter 8, Speci cations.
When the ambient temperature around the product is kept within the temperature range of the
operating environment speci cation (refer to \General" in Chapter 8), the product conforms to
the requirements of the safety standard. Furthermore, under that temperature environment, it
has been con rmed that the product still conforms to the requirements of the safety standard
when it is enclosed with cooling clearance as follows:
Upper

Instruction for Cleaning

For cleaning, wipe with soft cloth that is soaked with water and wrung tightly without undue
pressure.
1-8
Getting Started
Rear
Side
Conditions
250 mm
100 mm
15 mm
4349B