Agilent Technologies 4349B Manual page 128

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

SOURce Subsystem
COMMAND
SOURce
:VOLTage
[:LEVel]
:SOURce:VOLTage
<
>
numeric value
Entries or queries the external source voltage value, with which DUT's resistance is calculated
using measured current value.
< numeric value > is,
numeric
0 V to 5000 V
5000 V
0 V
Query response is < numeric value > .
Note
The 4349B does not have a test voltage source, and this command is not used
to control the test voltage level but to enter the voltage value which is used to
calculate resistance.
5-24
GPIB Reference
f
j
j
j
g
1
2
3
4
[:IMMediate]
[:AMPLitude]
f
j
j
j
g
1
2
3
4
[:LEVel][:IMMediate][:AMPLitude]
selects the channel.
PARAMETER
< numeric value >
4349B