4349B
:STATus:PRESet
Clears the OPERation and QUEStionable register groups. Both event and enable registers are
cleared. (no query)
:STATus:QUEStionable[:EVENt]?
Returns the contents of the event register of the standard questionable status group. (query
only)
Query response is the decimal expression of the contents of the register, in < NR1 > format.
The 4349B has no operation to report the event to the questionable status event register, all of
whose bits are always set to 0.
:STATus:QUEStionable:CONDition?
Returns the contents of the condition register of the standard questionable status group.
(Query only)
Query response is the decimal expression of the contents of the register, in < NR1 > format.
The 4349B has no operation to report the event to the questionable status condition register, all
of whose bits are always 0.
<
>
:STATus:QUEStionable:ENABle
numeric value
Sets or queries the enable register of standard questionable status group.
< numeric value > is,
numeric
decimal expression of the contents of the register
Query response is decimal expression of the contents of the register, in < NR1 > format.