DATA Subsystem
The DATA subsystem commands are used to store and read measurement data to and from the
4349B's data bu er. The 4349B provides a data bu er which is named
data up to 50 measurements into
COMMAND
DATA
[:DATA]?
:FEED
:POINts

:DATA[:DATA]? DBUF

Returns the data held in the data bu er,
commands in the FORMat subsystem. (query only)
This query must have the parameter
Query response is,
< set1 > , < set2 > , . . . < setn >
Where,
< set1 >
: data set of the rst measurement point
< set2 >
: data set of the second measurement point
. .
.
< setn >
: data set of the last measured point ( n is speci ed by
command)
Each data set consists of following data:
< stat1 > , < data1 > , < comp1 > , < stat2 > , < data2 > , < comp2 > , < stat3 > , < data3 > , < comp3 > ,
< stat4 > , < data4 > , < comp4 >
Where,
< stat1-4 > measured status of each channel
0 : normal
1 : overload
2 : No-Contact
< data1-4 > measured data of each channel
< comp1-4 > comparison result of each channel
0 : comparator OFF
1 : In
2 : High
4 : Low
8 : No-Contact
.
:CONTrol
, which is the name of the data bu er to be read.
PARAMETER
DBUF
DBUF, < data handle >
f
j
DBUF,
ALWays
NEVer
DBUF, < numeric value >
, according to the format determined by