The typical pull-up resistor values are:
8. Reinstall the A1 main board, the A2 board, and the cover.
B-10
Handler Interface Installation
Pull-up
Pull-up Resistor
Voltage
Agilent Part Number
5 V
0757-0278 (1.78 k )
9 V
0757-0279 (3.16 k )
12 V
0698-3154 (4.22 k )
15 V
0757-0438 (5.11 k )
EXT DCV2
EXT TRIG
SW1-1 SW1-2
5 to 6 V
Close
6 to 9 V
Close
9 to 15 V
Open
KEY LOCK
SW1-3
Close
Close
Open
Close
Close
Open