Ball should be within ring. (Figure 170)
4.1 Too Left
If the ball is too left then adjust arm at points in Figure 171.
Loosen #4 a quarter of a turn then loosen up #1 and #2.
Turn #3 toward the column and retighten #1 and #2. Also retighten #4 a
quarter of a turn.
Take another exposure. If aligned, re-check Beam check.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 170
Figure 171