Too Left - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

Ball should be within ring. (Figure 170)

4.1 Too Left

If the ball is too left then adjust arm at points in Figure 171.
Loosen #4 a quarter of a turn then loosen up #1 and #2.
Turn #3 toward the column and retighten #1 and #2. Also retighten #4 a
quarter of a turn.
Take another exposure. If aligned, re-check Beam check.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 170
Figure 171
- 81 -
81