Loosen the nut on the left and right locking screws. (Figure 181)
Loosen the right 3mm locking screw and tighten the 3mm left locking screw.
(Figure 182)
Take another exposure. If aligned tighten the right nut and tighten the two
screws on the bottom.
Move onto Beam Check.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 181
Figure 182