Elbow - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

3.1.b Elbow
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 100)
Figure 100
Put sensor coupling back onto spindle and re-tighten set screw. Re-
check still in the green and center.
When aligned take out the pins then press Done.
Unscrew the set screw with a 2mm allen to get the sensor holder off of
the cable guide. (Figure 101)
Figure 101
- 57 -
57