Pan Mode X-Collimator (61.1) - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

2.1 Pan mode X-Collimator (61.1) (Figure 27)
Now press the exposure button to see where the beam is hitting the screen.
The beam image should appear within the borders of the rectangle markings
on the screen. (Figure 28)
NOTE: The beam will need to be half of the distance of the box.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 27
U P
Figure 28
- 23 -
23