Image should come out as covering top to bottom on the fluorescent screens.
(Figure 141)
3.3.a Too High
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 141
Press the y-top limit down and y-bottom limit down. (Figure 142)
Press Test and take another exposure. If the beam is aligned then press
Done and Move onto the Secondary collimator (beaver tail).
- 77 -
Figure 142
77