Place the Ball Phantom into the patient positioning mechanism with the Frankfort
plane alignment tool on top of the Ball Phantom. (Figure 106)
NOTE: Make sure the layer light is on the line of the ball phantom then zero out the layer light.
Press i then press Technical Calibration (60.0) (Figure 107)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 106
Figure 107