Too Low; Ceph Head Leveling - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Hello, I am your AI assistant. How can I help you?
Table of Contents

Advertisement

3.3.b Too Low
4

Ceph Head Leveling

Put the left and right ear posts into their respective holders. (Figure 144)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Press the y-top limit up and y-bottom limit up. (Figure 143)
Press Test and take another exposure. If the beam is aligned then press
Done. Move onto the Secondary collimator (beaver tail).
Figure 144
- 78 -
Figure 143
78

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ProMax Pan/Ceph and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF