3.3.b Too Low
4
Ceph Head Leveling
Put the left and right ear posts into their respective holders. (Figure 144)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Press the y-top limit up and y-bottom limit up. (Figure 143)
Press Test and take another exposure. If the beam is aligned then press
Done. Move onto the Secondary collimator (beaver tail).
Figure 144