Too Low; Angled - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

2.2.b Too Low
2.2.c Angled
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Press Test and take another exposure.
Press the Y-pan collimator bottom limit button up. (Figure 35)
Figure 35
Press Test and take another exposure.
If aligned then press Done and move onto Ball Phantom.
Press the i button in the lower left hand corner. (Figure 36)
Figure 36
- 27 -