4.3 Planmeca Proline XC pan/ceph X-ray - checking the alignment of the radiation
beam
Panoramic X-ray beam
Cassette drive
assembly
Secondary collimator
Installation manual
Cover
Beam
alignment
tool
Primary
collimator
Remove the inner covers from tube head assembly and
the cassette drive assembly. Four screws hold the tube
head cover in position, these are located at the rear of
the tube head assembly. The screws that hold the cas-
sette drive cover in place are located at the front and
lower side of the cover. The primary collimator mecha-
nism is located under the tube head cover.
Tube head cover
Place the beam alignment tool into the secondary colli-
mator, it is magnetic so it will stay in position. Note which
way round it goes. Leave the cassette in position as it will
reduce the amount of scattered radiation.
Planmeca Proline XC X-ray unit 21
CHECKING THE ALIGNMENT