Planmeca ProMax Pan/Ceph Calibration Manual page 99

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

7.3.a Not together
Figure 205a
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
If the image comes out separated check to make sure that the hatchet is
placed in the holder correctly. Re-do Beam Check.
7.3.a.i
Raising up
If hatchet is place correctly, then raise the beaver tail
up.
Loosen the two screws on the inside of the beaver tail
and tighten the vertical screw on the outside of the
beaver tail. (Figure 205a-b)
- 99 -
Figure 205b
99