4 Patient Positioning Lights Alignment

After aligning the Ball Phantom, turn the patient positioning lights on by pressing the
temple support button. (Figure 105)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 104)
Figure 104
When aligned take out the pins then press Done.
Figure 105