Too Left/Right - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

3.2.a Too Left/Right
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
If the beam is too left, move the x-right limit to the right. (Figure 137)
Figure 137
Press Test and take another exposure.
If the beam is too right, move the x-right limit to the left. (Figure 138)
Figure 138
Press Test and take another exposure. If the beam is aligned then press
Done and move onto the Ceph y-collimator.