Secondary Collimator (Beaver Tail)
Unscrew the three 3mm allen screws. (Figure 176a-b)
Figure 176a
Attach the collimator and screw in the three 3mm allens. (Figure 177a-b)
Figure 177a
Press i then Technical calibration then Primary Collimator and select "Ceph mode
beam position" (61.6)
Press Test and take an exposure. (Figure 178)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 176b
Figure 177b
Figure 178
- 84 -