Take an exposure. The image should be the same on the top, bottom, left, and
right. (Figure 199)
7
Calibrating
Press Exit on the touch screen, if just finished doing a Beam Check.
7.1 kV / mA for Cephalometric
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 199
Binnings
kV
mA
4x4
60
4
3x3
60
7