Too Low - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

Loosen the left 3mm locking screw and tighten the 3mm right locking screw.
(Figure 185)
Take another exposure. If aligned tighten the left nut and tighten the two
screws on the bottom.
Move onto Beam Check.

5.3 Too Low

Loosen the two 3mm allen screws on the inside of the collimator. (Figure 186)