Angled Left/Right; Left - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

The collimator will move to expose one screw then push the lower blade up
and expose the second screw. (Figure 41)
NOTE: This allows the two screws that adjust the angle to be visible.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 41
2.2.c.i Angled Left/Right
Loosen the two hidden screws with a 2.5 mm allen at # 1.
(Figure 42)
Figure 42
Twist collimator to the right if angled Right to Left
Twist collimator to the left if angled Left to Right
Tighten down the screws and take another exposure.
Go back into Pan X-Collimator and Y-Collimator and press Test in
both allowing the collimator to go to the home position.
Take another exposure. Press Done until you get back to the Technical
Calibration Screen.
- 30 -
30

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents