A.ii Too Wide; Pan Mode Y-Collimator (61.2); Too High - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

2.2 Pan mode Y Collimator (61.2)
Now check the top and bottom of the beam.
2.2.a Too High
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018

2.1.a.ii Too Wide

Press the X-pan collimator width to the left (Figure 33)
Figure 33
Press Test and take another exposure.
If aligned then press Done and move onto the Y
Collimator.
Press the Y-pan collimator top limit button down. (Figure 34)
Figure 34
- 26 -
26