Planmeca ProMax Pan/Ceph Calibration Manual page 91

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

5.5.b Right
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
If angled right, then loosen the nuts and loosen the left locking screw
and tighten the right locking screw. (Figure 191)
Figure 192
Take another exposure. If aligned then re-tighten the nuts and move
onto Beam Check.
- 91 -
91