4.2 Frankfort Light Beam

If the Frankfort light is not straight on the Frankfort plane tool, then adjust the
laser. (See Figure 109)

4.3 Midsagittal Light Beam

If the Midsagittal light is not lined up with the center line on the Ball
phantom, then finger adjust the laser until it lines up. (See Figure 109)