A.ii Too Fat - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

NOTE: The new shoulder number should be higher than older number.
Press Lock and take the pin out and then press Done. Take another exposure
and check ball size again. If not repeat until correct size.

2.1.a.ii Too Fat

If the ball is too fat (Figure 76) then adjust in Pots.
Press i then select Technical Calibration (60.0) then select Position Sensor
Calibration. (64.0). (Figure 77)
Press Position Sensor (64.1) (Figure 78)
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018
Figure 76
Figure 77
Figure 78
- 46 -
46