Too Thin - Planmeca ProMax Pan/Ceph Calibration Manual

Panoramic with cephalometric x-ray unit
Table of Contents

Advertisement

2.1.a.i
Printed copies of this document are considered uncontrolled.
22811.20.Rev001
07.31.2018

Too thin

If the ball is too thin (Figure 70) then adjust worm gear in Pots.
Figure 70
Press i then select Technical Calibration (60.0) then select Position
Sensor Calibration. (64.0). (Figure 71)
Figure 71
Press Position Sensor (64.1) (Figure 72)
Figure 72
Press Check and write down what the shoulder position number is.
- 44 -
44