Integrate Group - YOKOGAWA WT3001E User Manual

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

6.15 INTEGrate Group

The commands in this group deal with integration.
You can make the same settings and inquiries as when INTEG on the front panel is used.
:INTEGrate?
Function
Queries all settings related to the integration.
:INTEGrate?
Syntax
:INTEGRATE? -> :INTEGRATE:
Example
MODE NORMAL;ACAL 0;TIMER 1,0,0
:INTEGrate:ACAL
Function
Turns ON/OFF the auto calibration or queries the
current setting.
:INTEGrate:ACAL {<Boolean>}
Syntax
:INTEGrate:ACAL?
:INTEGRATE:ACAL OFF
Example
:INTEGRATE:ACAL? ->
:INTEGRATE:ACAL 0
:INTEGrate:MODE
Function
Sets the integration mode or queries the current
setting.
:INTEGrate:MODE {NORMal|CONTinuous|
Syntax
RNORmal|RCONtinuous}
:INTEGrate:MODE?
NORMal = Normal integration mode
CONTinuous = Continuous integration mode
RNORmal = Real-time normal integration mode
RCONtinuous = Real-time continuous integration
mode
:INTEGRATE:MODE NORMAL
Example
:INTEGRATE:MODE? ->
:INTEGRATE:MODE NORMAL
:INTEGrate:RESet
Function
Resets the integrated value.
:INTEGrate:RESet
Syntax
:INTEGRATE:RESET
Example
:INTEGrate:RTIMe?
Function
Queries the integration start and stop times for
real-time integration mode.
:INTEGrate:RTIMe<x>?
Syntax
:INTEGRATE:RTIME? ->
Example
:INTEGRATE:RTIME:
START 2005,1,1,0,0,0;
END 2005,1,1,1,0,0
6-74
:INTEGrate:RTIMe:{STARt|END}
Function
Sets the integration {start|stop} time for real-time
integration mode or queries the current setting.
:INTEGrate:RTIMe:{STARt|
Syntax
END} {<NRf>,<NRf>,<NRf>,<NRf>,
<NRf>,<NRf>}
:INTEGrate:RTIMe:{STARt|END}?
{<NRf>, <NRf>, <NRf>, <NRf>, <NRf>, <NRf>} =
2001, 1, 1, 0, 0, 0 to 2099, 12, 31, 23, 59, 59
1st <NRf> = 2001 to 2099 (year)
2nd <NRf> = 1 to 12
3rd <NRf> = 1 to 31
4th <NRf> = 0 to 23
5th <NRf> = 0 to 59
6th <NRf> = 0 to 59
:INTEGRATE:RTIME:
Example
START 2005,1,1,0,0,0
:INTEGRATE:RTIME:START? ->
:INTEGRATE:RTIME:
START 2005,1,1,0,0,0
:INTEGrate:STARt
Function
Starts integration.
:INTEGrate:STARt
Syntax
:INTEGRATE:START
Example
:INTEGrate:STATe?
Function
Queries the integration condition.
:INTEGrate:STATe?
Syntax
:INTEGRATE:STATE? -> RESET
Example
Description The response information is as follows:
RESet = Integration reset
READy = Waiting (real-time integration mode)
STARt = Integration in progress
STOP = Integration stop
ERRor = Abnormal integration termination
(integration overflow, power failure)
TIMeup = Integration stop due to integration timer
time
:INTEGrate:STOP
Function
Stops integration.
:INTEGrate:STOP
Syntax
:INTEGRATE:STOP
Example
(month)
(day)
(hour)
(minute)
(second)