6.15 INTEGrate Group

The commands in this group deal with integration.
You can make the same settings and inquiries as when INTEG on the front panel is used.
:INTEGrate?
Function
Queries all settings related to the integration.
Syntax
:INTEGrate?
Example
:INTEGRATE? -> :INTEGRATE:
MODE NORMAL;ACAL 0;TIMER 1,0,0
:INTEGrate:ACAL
Function
Turns ON/OFF the auto calibration or queries
the current setting.
Syntax
:INTEGrate:ACAL {<Boolean>}
:INTEGrate:ACAL?
Example
:INTEGRATE:ACAL OFF
:INTEGRATE:ACAL? ->
:INTEGRATE:ACAL 0
:INTEGrate:MODE
Function
Sets the integration mode or queries the current
setting.
Syntax
:INTEGrate:MODE {NORMal|CONTinuous|
RNORmal|RCONtinuous}
:INTEGrate:MODE?
NORMal = Normal integration mode
CONTinuous = Continuous integration mode
RNORmal = Real-time normal integration mode
RCONtinuous = Real-time continuous
integration mode
Example
:INTEGRATE:MODE NORMAL
:INTEGRATE:MODE? ->
:INTEGRATE:MODE NORMAL
:INTEGrate:RESet
Function
Resets the integrated value.
Syntax
:INTEGrate:RESet
Example
:INTEGRATE:RESET
:INTEGrate:RTIMe?
Function
Queries the integration start and stop times for
real-time integration mode.
Syntax
:INTEGrate:RTIMe<x>?
Example
:INTEGRATE:RTIME? ->
:INTEGRATE:RTIME:
START 2005,1,1,0,0,0;
END 2005,1,1,1,0,0
6-74
:INTEGrate:RTIMe:{STARt|END}
Function
Sets the integration {start|stop} time for real-
time integration mode or queries the current
setting.
Syntax
:INTEGrate:RTIMe:{STARt|
END} {<NRf>,<NRf>,<NRf>,<NRf>,
<NRf>,<NRf>}
:INTEGrate:RTIMe:{STARt|END}?
{<NRf>, <NRf>, <NRf>, <NRf>, <NRf>, <NRf>}
= 2001, 1, 1, 0, 0, 0 to 2099, 12, 31, 23, 59, 59
1st <NRf> = 2001 to 2099 (year)
2nd <NRf> = 1 to 12
3rd <NRf> = 1 to 31
4th <NRf> = 0 to 23
5th <NRf> = 0 to 59
6th <NRf> = 0 to 59
Example
:INTEGRATE:RTIME:
START 2005,1,1,0,0,0
:INTEGRATE:RTIME:START? ->
:INTEGRATE:RTIME:
START 2005,1,1,0,0,0
:INTEGrate:STARt
Function
Starts integration.
Syntax
:INTEGrate:STARt
Example
:INTEGRATE:START
:INTEGrate:STATe?
Function
Queries the integration condition.
Syntax
:INTEGrate:STATe?
Example
:INTEGRATE:STATE? -> RESET
Description The response information is as follows:
RESet = Integration reset
READy = Waiting (real-time integration mode)
STARt = Integration in progress
STOP = Integration stop
ERRor = Abnormal integration termination
(integration overflow, power failure)
TIMeup = Integration stop due to integration
timer time
:INTEGrate:STOP
Function
Stops integration.
Syntax
:INTEGrate:STOP
Example
:INTEGRATE:STOP
(month)
(day)
(hour)
(minute)
(second)