YOKOGAWA WT3001E User Manual page 83

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

:DISPlay:NUMeric[:NORMal]:{VAL4|VAL8
|VAL16}:PRESet
Function
Sets the displayed items on the numeric display
({4-value|8-value|16-value} display) to a preset
pattern.
:DISPlay:NUMeric[:NORMal]:{VAL4|
Syntax
VAL8|VAL16}:PRESet {<NRf>}
<NRf> = 1 to 4
:DISPLAY:NUMERIC:NORMAL:VAL4:
Example
PRESET 1
Description Regardless of what value (1 to 4) is specified
for <NRf>, the display pattern (order) of the
numeric display items will be the same as the
display order when Reset Items Exec of the ITEM
setting menu, which is displayed on the screen,
is executed. For details on the order of displayed
items when reset is executed, see the User's
Manual IM WT3001E-01EN.
:DISPlay:TRENd?
Function
Queries all settings related to the trend.
:DISPlay:TRENd?
Syntax
:DISPLAY:TREND? -> :DISPLAY:TREND:
Example
FORMAT SINGLE;T1 1;T2 1;T3 1;T4 1;
T5 1;T6 1;T7 1;T8 1;T9 0;T10 0;
T11 0;T12 0;T13 0;T14 0;T15 0;
T16 0;TDIV 0,0,3;ITEM1:
FUNCTION U,1,TOTAL;SCALING:
MODE AUTO;
VALUE 100.0E+00,-100.0E+00;:
DISPLAY:TREND:ITEM2:
FUNCTION I,1,TOTAL;SCALING:
MODE AUTO;
VALUE 100.0E+00,-100.0E+00;...
(omitted)...;:DISPLAY:TREND:NORMAL:
ITEM16:FUNCTION AH,1;SCALING:
MODE AUTO;
VALUE 100.0E+00,-100.0E+00
:DISPlay:TRENd:ALL
Function
Collectively turns ON/OFF all trends.
:DISPlay:TRENd:ALL {<Boolean>}
Syntax
:DISPLAY:TREND:ALL ON
Example
:DISPlay:TRENd:CLEar
Function
Clears the trend.
:DISPlay:TRENd:CLEar
Syntax
:DISPLAY:TREND:CLEAR
Example
IM WT3001E-17EN
:DISPlay:TRENd:FORMat
Function
Sets the display format of the trend or queries the
current setting.
:DISPlay:TRENd:FORMat {SINGle|DUAL|
Syntax
TRIad|QUAD}
:DISPlay:TRENd:FORMat?
:DISPLAY:TREND:FORMAT SINGLE
Example
:DISPLAY:TREND:FORMAT? ->
:DISPLAY:TREND:FORMAT SINGLE
:DISPlay:TRENd:ITEM<x>?
Function
Queries all settings related to the trend.
:DISPlay:TRENd:ITEM<x>?
Syntax
<x> = 1 to 16 (item number)
:DISPLAY:TREND:ITEM1? -> :DISPLAY:
Example
TREND:ITEM1:FUNCTION U,1,TOTAL;
SCALING:MODE AUTO;
VALUE 100.0E+00,-100.0E+00
:DISPlay:TRENd:ITEM<x>[:FUNCtion]
Function
Sets the trend item (function, element, and
harmonic order) or queries the current setting.
:DISPlay:TRENd:ITEM<x>
Syntax
[:FUNCtion] {NONE|<Function>,
<Element>[,<Order>]}
:DISPlay:TRENd:ITEM<x>:FUNCtion?
<x> = 1 to 16 (item number)
NONE = No display item
<Function> = {U|I|P|S|Q|...} (See the
function selection list (1) on page 6-44.)
<Element> = {<NRf>|SIGMa|SIGMB} (<NRf>
= 1 to 4)
<Order> = {TOTal|DC|<NRf>} (<NRf> = 1 to
100)
:DISPLAY:TREND:ITEM1:
Example
FUNCTION U,1,TOTAL
:DISPLAY:TREND:ITEM1:FUNCTION? ->
:DISPLAY:TREND:ITEM1:
FUNCTION U,1,TOTAL
Description • If <Element> is omitted, element 1 is set.
• If <Order> is omitted, TOTal is set.
• <Element> or <Order> is omitted from
response to functions that do not need them.
:DISPlay:TRENd:ITEM<x>:SCALing?
Function
Queries all settings related to the scaling of the
trend.
:DISPlay:TRENd:ITEM<x>:SCALing?
Syntax
<x> = 1 to 16 (item number)
:DISPLAY:TREND:ITEM1:SCALING? ->
Example
:DISPLAY:TREND:ITEM1:SCALING:
MODE AUTO;
VALUE 100.0E+00,-100.0E+00
6.7 DISPlay Group
1
2
3
4
5
6
7
App
Index
6-39

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt3002eWt3003eWt3004e