6.17 MOTor Group
The commands in this group deal with the motor evaluation function.
You can make the same settings and inquiries as when MOTOR SET (SHIFT+SCALING) on the front panel is used.
However, the commands in this group are valid only on models with the motor evaluation function (/MTR option).
:MOTor?
Function
Queries all settings related to the motor
evaluation function.
:MOTor?
Syntax
:MOTOR? -> :MOTOR:SPEED:
Example
TYPE ANALOG;RANGE 20.0E+00;AUTO 0;
SCALING 1.0000;UNIT "rpm";:MOTOR:
TORQUE:TYPE ANALOG;RANGE 20.0E+00;
AUTO 0;SCALING 1.0000;UNIT "Nm";:
MOTOR:PM:SCALING 1.0000;UNIT "W";:
MOTOR:FILTER:LINE OFF;:MOTOR:
SYNCHRONIZE NONE;POLE 2;SSPEED I1
:MOTor:FILTer?
Function
Queries all settings related to the input filter.
:MOTor:FILTer?
Syntax
:MOTOR:FILTER? -> :MOTOR:FILTER:LINE
Example
OFF
:MOTor:FILTer[:LINE]
Function
Sets the line filter or queries the current setting.
:MOTor:FILTer[:LINE]
Syntax
{OFF|<Frequency>}
:MOTor:FILTer:LINE?
OFF = Line filter OFF
<Frequency> = 100 Hz, 50 kHz (line filter ON,
cutoff frequency)
:MOTOR:FILTER:LINE OFF
Example
:MOTOR:FILTER:LINE? ->
:MOTOR:FILTER:LINE OFF
:MOTor:PM?
Function
Queries all settings related to the motor output
(Pm).
:MOTor:PM?
Syntax
:MOTOR:PM? -> :MOTOR:PM:
Example
SCALING 1.0000;UNIT "W"
:MOTor:PM:SCALing
Function
Sets the scaling factor used for motor output
computation or queries the current setting.
:MOTor:PM:SCALing {<NRf>}
Syntax
:MOTor:PM:SCALing?
<NRf> = 0.0001 to 99999.9999
:MOTOR:PM:SCALING 1
Example
:MOTOR:PM:SCALING? ->
:MOTOR:PM:SCALING 1.0000
IM WT3001E-17EN
:MOTor:PM:UNIT
Function
Sets the unit to add to the motor output
computation result or queries the current setting.
:MOTor:PM:UNIT {<String>}
Syntax
:MOTor:PM:UNIT?
<String> = Up to 8 characters
:MOTOR:PM:UNIT "W"
Example
:MOTOR:PM:UNIT? ->
:MOTOR:PM:UNIT "W"
Description This command does not affect the computation
result.
:MOTor:POLE
Function
Sets the motor's number of poles or queries the
current setting.
:MOTor:POLE {<NRf>}
Syntax
:MOTor:POLE?
<NRf> = 1 to 99
:MOTOR:POLE 2
Example
:MOTOR:POLE? -> :MOTOR:POLE 2
:MOTor:SPEed?
Function
Queries all settings related to the rotating speed.
:MOTor:SPEed?
Syntax
:MOTOR:SPEED? -> :MOTOR:SPEED:
Example
TYPE ANALOG;RANGE 20.0E+00;AUTO 0;
SCALING 1.0000;UNIT "rpm"
:MOTor:SPEed:AUTO
Function
Turns ON/OFF the voltage auto range of the
revolution signal input (analog input format) or
queries the current setting.
:MOTor:SPEed:AUTO {<Boolean>}
Syntax
:MOTor:SPEed:AUTO?
:MOTOR:SPEED:AUTO ON
Example
:MOTOR:SPEED:AUTO? ->
:MOTOR:SPEED:AUTO 1
Description This command is valid when the revolution signal
input type ( :MOTor:SPEed:TYPE ) is "ANALog
(analog input)."
1
2
3
4
5
6
7
App
Index
6-81