5.14 INTEGrate Group

The commands in this group deal with integration.
You can make the same settings and queries that you can make by pressing INTEG on the front panel.
:INTEGrate?
Function
Queries all integration settings.
:INTEGrate?
Syntax
:INTEGrate:ACAL
Function
Sets or queries the on/off status of integration
auto calibration.
:INTEGrate:ACAL {<Boolean>}
Syntax
:INTEGrate:ACAL?
:INTEGRATE:ACAL OFF
Example
:INTEGRATE:ACAL? ->
:INTEGRATE:ACAL 0
:INTEGrate:INDependent
Function
Sets or queries the on/off status of independent
element integration.
:INTEGrate:INDependent {<Boolean>}
Syntax
:INTEGrate:INDependent?
:INTEGRATE:INDEPENDENT OFF
Example
:INTEGRATE:INDEPENDENT? ->
:INTEGRATE:INDEPENDENT 0
:INTEGrate:MODE
Function
Sets or queries the integration mode.
:INTEGrate:MODE {NORMal|CONTinuous|
Syntax
RNORmal|RCONtinuous}
:INTEGrate:MODE?
NORMal = Normal integration mode
CONTinuous = Continuous integration
mode
RNORmal = Real-time normal
integration mode
RCONtinuous = Real-time continuous
integration mode
:INTEGRATE:MODE NORMAL
Example
:INTEGRATE:MODE? ->
:INTEGRATE:MODE NORMAL
:INTEGrate:QMODe?
Function
Queries the electric current modes for electric
current integration of all elements.
:INTEGrate:QMODe?
Syntax
:INTEGrate:QMODe[:ALL]
Function
Collectively sets the electric current mode for
electric current integration of all elements.
:INTEGrate:QMODe[:ALL]
Syntax
{RMS|MEAN|DC|RMEAN|AC}
:INTEGRATE:QMODE:ALL DC
Example
IM WT1801-17EN
:INTEGrate:QMODe:ELEMent<x>
Function
Sets or queries the electric current mode for
electric current integration of the specified
element.
:INTEGrate:QMODe:ELEMent<x>
Syntax
{RMS|MEAN|DC|RMEAN|AC}
:INTEGrate:QMODe:ELEMent<x>?
<x> = 1 to 6 (element)
:INTEGRATE:QMODE:ELEMENT1 DC
Example
:INTEGRATE:QMODE:ELEMENT1? ->
:INTEGRATE:QMODE:ELEMENT1 DC
Description Regardless of the independent element
integration setting (:INTEGrate:INDependent),
the WT1800 operates according to the electric
current mode of the specified element.
:INTEGrate:RESet
Function
Resets the integrated value.
:INTEGrate:RESet {[<NRf>][,<NRf>]
Syntax
[,<NRf>][,<NRf>][,<NRf>][,<NRf>]}
<NRf> = 1 to 6 (element)
:INTEGRATE:RESET (resets all
Example
elements)
:INTEGRATE:RESET 1,2,3 (resets the
specified elements)
Description • When independent element integration (:
INTEGrate:INDependent) is set to ON (1),
you can use the parameters to specify which
elements to perform the operation on. If you
omit the parameters, the operation will be
performed on all the elements.
• When independent element integration (:
INTEGrate:INDependent) is set to OFF (0), you
cannot use the parameters to specify elements.
:INTEGrate:RTALl:{STARt|END}
Function
Collectively sets the integration start or end time
of all elements for real-time integration mode.
:INTEGrate:RTALl:{STARt|END} {<NRf>,
Syntax
<NRf>,<NRf>,<NRf>,<NRf>,<NRf>}
{<NRf>,<NRf>,<NRf>,<NRf>,
<NRf>,<NRf>} = 2001,1,1,0,0,0 to
2099,12,31,23,59,59
First <NRf> = 2001 to 2099 (year)
Second <NRf> = 1 to 12 (month)
Third <NRf> = 1 to 31 (day)
Fourth <NRf> = 0 to 23 (hour)
Fifth <NRf> = 0 to 59 (minute)
Sixth <NRf> = 0 to 59 (second)
:INTEGRATE:RTALL:START
Example
2010,1,1,0,0,0