• For IEC61000-3-3 Ed3.0, "TMAX" is returned
as a response to a setting query.
:NUMeric:FLICker:FUNCtion:NUMber
Function
Sets the number of measured flicker data output
by ":NUMeric:FLICker:FUNCtion:VALue?" or
queries the current setting.
:NUMeric:FLICker:FUNCtion:
Syntax
NUMber {<NRf>|ALL}
:NUMeric:FLICker:FUNCtion:NUMber?
<NRf> = 1 to 32(ALL)
:NUMERIC:FLICKER:FUNCTION:NUMBER 8
Example
:NUMERIC:FLICKER:FUNCTION:NUMBER ->
:NUMERIC:FLICKER:FUNCTION:NUMBER 8
Description • Only available with the flicker measurement
function (/FL option).
• If parameters are omitted from " :NUMeric:F
LICker:FUNCtion:VALue? ", numeric data
from 1 to (specified value) is output in order.
• The initial setting for the number of numeric
data is 8.
:NUMeric:FLICker:FUNCtion:VALue?
Function
Queries the measured flicker data (variable
format).
:NUMeric:FLICker:FUNCtion:
Syntax
VALue? {<NRf>}
:NUMeric:FLICker:FUNCtion:VALue?
<NRf> = 1 to 32 (item number)
Example
• Example when <NRf> is specified
:NUMERIC:FLICKER:FUNCTION:
VALUE? 4 -> 1.52E+00
• Example when <NRf> is omitted
:NUMERIC:FLICKER:FUNCTION:
VALUE? -> 600,229.75E+00,
50.000E+00,1.52E+00,1.56E+00,
3E+00,0.43E+00,0.17E+00
• Example in which ":NUMeric:FORMat" is set to
" FLOat ".
:NUMERIC:FLICKER:FUNCTION:VALUE?
-> #4 (number of bytes, four digits) (data
byte sequence)
Description • Only available with the flicker measurement
function (/FL option).
• When <NRf> is specified, only the numeric
data for that item is output.
IM WT3001E-17EN
• If <NRf> is omitted, numeric data from the item
number in " :NUMeric:FLICker:FUNCtion:
NUMber " is output in order.
• The format of individual numeric data that is
output is as follows:
(1) Data when normal
• Elapsed measurement time (TIME)
ASCII: <NR1> format in units of seconds
(Example :for 1 hour (1:00:00), 3600)
FLOAT: IEEE single-precision floating
point (4-byte) format in units of seconds
(Example :for 1 hour (1:00:00), 0x45610000)
• No items (NONE)
ASCII: "NAN" (Not A Number)
FLOAT: 0x7E951BEE (9.91E+37)
• Other than above
ASCII: <NR3> format (mantissa, 5 digits;
exponent, 2 digits, Example :229.87E+00)
FLOAT: IEEE single-precision floating
point (4-byte) format
(2) Error Data
• Data does not exist (display: "-----")
ASCII: "NAN" (Not A Number)
FLOAT: 0x7E951BEE (9.91E+37)
• Overrange (display: "-O-L-")
• Overflow (display: "-O-F-")
• Data over (display: "Error")
• No steady-state condition (display: "Undef")
ASCII: "INF" (INFinity)
FLOAT: 0x7E94F56A (9.9E+37)
:NUMeric:FLICker:INFOrmation?
Function
Queries all settings related to output of flicker
judgment results (variable format).
:NUMeric:FLICker:INFOrmation?
Syntax
:NUMERIC:FLICKER:INFORMATION? ->
Example
:NUMERIC:FLICKER:INFORMATION:
NUMBER 7;ITEM1 JTOTAL,TOTAL;
ITEM2 JDC,1,ALL;ITEM3 JDMAX,1,ALL;
ITEM4 JDT,1,ALL;ITEM5 JPST,1,ALL;
ITEM6 JPLT,1;ITEM7 JTOTAL,1
Description • Only available with the flicker measurement
function (/FL option).
• For the values of
" :NUMeric:FLICker:INFOrmation:
ITEM<x> ", only the number of
numeric data output items specified in
" :NUMeric:FLICker:FUNCtion:
INFOrmation:NUMber " are output.