Flicker Group - YOKOGAWA WT3001E User Manual

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

6.9

FLICker Group

The FLICker group contains commands related to flicker measurement.
These commands allow you to enter and query the same settings that are available under ITEM in the Flicker Items
menu and under FORM in the Flicker Form menu on the front panel.
Note that the commands in this group are only available with the flicker measurement function (/FL option).
:FLICker?
Function
Queries all settings related to flicker
measurement.
:FLICker?
Syntax
:FLICKER? -> :FLICKER:
Example
MEASUREMENT FLICKER;ELEMENT1 1;
ELEMENT2 0;ELEMENT3 0;
INTERVAL 10,0;COUNT 12;
FREQUENCY 50;UN:MODE AUTO;
VALUE 230.00;:FLICKER:DC:STATE 1;
LIMIT 3.30;:FLICKER:DMAX:STATE 1;
LIMIT 4.00;:FLICKER:DT:STATE 1;
LIMIT 500,3.30;:FLICKER:PST:
STATE 1;LIMIT 1.00;:FLICKER:PLT:
STATE 1;LIMIT 0.65;NVALUE 12;:
FLICKER:DMIN:LIMIT 0.10
:FLICker:COUNt
Function
Sets the number of measurements for the short-
term flicker value Pst or queries the current
setting.
:FLICker:COUNt {<NRf>}
Syntax
:FLICker:COUNt?
<NRf> = 1 to 99 (no. of measurements)
:FLICKER:COUNT 12
Example
:FLICKER:COUNT? ->
:FLICKER:COUNT 12
Description This setting is available with normal flicker
measurement ( :FLICker:MEASurement
FLICker ).
For measurement of dmax caused by manual
switching ( :FLICker:MEASurement DMAX ), the
number of measurements is fixed at 24.
:FLICker:DC?
Function
Sets all settings related to the relative steady-
state voltage change dc or queries the current
setting.
:FLICker:DC?
Syntax
:FLICKER:DC? ->
Example
:FLICKER:DC:STATE 1;LIMIT 3.30
6-52
:FLICker:DC:LIMit
Function
Sets the limit of the relative steady-state voltage
change dc or queries the current setting.
:FLICker:DC:LIMit {<NRf>}
Syntax
:FLICker:DC:LIMit?
<NRf> = 1.00 to 99.99 (limit[%])
:FLICKER:DC:LIMIT 3.30
Example
:FLICKER:DC:LIMIT? ->
:FLICKER:DC:LIMIT 3.30
:FLICker:DC[:STATe]
Function
Turns ON/OFF judgment of the relative steady-
state voltage change dc or queries the current
setting.
:FLICker:DC[:STATe] {<Boolean>}
Syntax
:FLICker:DC:STATe?
:FLICKER:DC:STATE ON
Example
:FLICKER:DC:STATE? ->
:FLICKER:DC:STATE 1
:FLICker:DISPlay?
Function
Queries all settings related to flicker measurement
display.
:FLICker:DISPlay?
Syntax
:FLICKER:DISPLAY? ->
Example
:FLICKER:DISPLAY:ELEMENT 1;PERIOD 1
:FLICker:DISPlay:ELEMent
Function
Sets the element to be displayed for flicker
measurement display or queries the current
setting.
:FLICker:DISPlay:ELEMent {<NRf>}
Syntax
:FLICker:DISPlay:ELEMent?
<NRf> = 1 to 4 (element)
:FLICKER:DISPLAY:ELEMENT 1
Example
:FLICKER:DISPLAY:ELEMENT? ->
:FLICKER:DISPLAY:ELEMENT 1
Description You can make the same setting or query with the
" :DISPlay:FLICker:ELEMent " command.
IM WT3001E-17EN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt3002eWt3003eWt3004e

Table of Contents