5.5 Synchronization with the Controller
Using the Extended Event Register
The changes in the condition register can be reflected
in the extended event register (page 7-7).
Example :STATus:FILTer1 FALL;:STATus:
EESE 1;EESR?;*SRE 8<PMT>
(Read the response to STATus:EESR? )
Loop
(Wait for a service request)
:NUMeric[:NORMal]:VALue?<PMT>
(Read the response to
:NUMeric[:NORMal]:VALue? )
:STATus:EESR?<PMT>
(Read the response to :STATus:EESR? )
(Return to LOOP )
The " STATus:FILTer1 FALL " command
sets the transition filter so that bit 0
( FILTer1 ) of the extended event register is
set to 1 when bit 0 of the condition register
changes from 1 to 0 .
The " :STATus:EESE 1 " command is used
to reflect only bit 0 of the extended event
register to the status byte.
The " STATus:EESR? " command is used to
clear the extended event register.
The " *SRE 8 " command is used to generate
a service request solely on the cause of the
extended event register.
The " :NUMeric[:NORMal]:VALue? "
command is not executed until a service
request is generated.
5-10
Using the COMMunicate:WAIT command
The " COMMunicate:WAIT " command halts
communications until a specific event is generated.
Example :STATus:FILTer1 FALL;:STATus:
EESR?<PMT>
(Read the response to STATus:EESR? )
Loop
COMMunicate:WAIT 1<PMT>
:NUMeric[:NORMal]:VALue?<PMT>
(Read the response to
:NUMeric[:NORMal]:VALue? )
:STATus:EESR?<PMT>
(Read the response to :STATus:EESR? )
(Return to LOOP )
For a description of " STATus:
FILTer1 FALL " and " STATus:EESR? "
see the previous section regarding
the extended event register.
The " COMMunicate:WAIT 1 " command
indicates that the program will wait for bit
0 of the extended event register to be set
to " 1 ."
The " :NUMeric[:NORMal]:VALue? "
command is not executed until bit 0 of
the extended event register is set to " 1 ."