6.16 MEASure Group
:MEASure:PC:IEC
Function
Sets the equation used to compute Pc (Corrected
Power) or queries the current setting.
:MEASure:PC:IEC {<NRf>}
Syntax
:MEASure:PC:IEC?
<NRf> = 1976, 1993 or 2011
:MEASURE:PC:IEC 1976
Example
:MEASURE:PC:IEC? ->
:MEASURE:PC:IEC 1976
Description • Specify the year when the equation used to
calculate the Pc was issued by IEC76-1.
• If 2011 is specified, "IEC 1993" is returned as a
response to a setting query.
:MEASure:PC:P<x>
Function
Sets the parameter used to compute Pc
(Corrected Power) or queries the current setting.
:MEASure:PC:P<x> {<NRf>}
Syntax
:MEASure:PC:P<x>?
<x> = 1, 2 (P1, P2)
<NRf> = 0.0001 to 9.9999
:MEASURE:PC:P1 0.5
Example
:MEASURE:PC:P1? ->
:MEASURE:PC:P1 0.5000
Description This parameter is used when the
" :MEASure:PC:IEC " setting is set to
"1976(IEC76-1(1976))."
:MEASure:PHASe
Function
Sets the display format of the phase difference or
queries the current setting.
:MEASure:PHASe {<NRf>}
Syntax
:MEASure:PHASe?
<NRf> = 180 or 360
:MEASURE:PHASE 180
Example
:MEASURE:PHASE? ->
:MEASURE:PHASE 180
Description Displays the phase using ±0 to 180° (Lead/Lag)
for "180" and 0 to 360° for "360."
:MEASure:SAMPling
Function
Sets the sampling frequency or queries the
current setting.
:MEASure:SAMPling {AUTO|CLKA|CLKB|
Syntax
CLKC}
:MEASure:SAMPling?
:MEASURE:SAMPLING AUTO
Example
:MEASURE:SAMPLING? ->
:MEASURE:SAMPLING AUTO
Description For details on the sampling frequency
corresponding to {AUTO|CLKA|CLKB|CLKC} ,
see the User's Manual IM WT3001E-01EN.
6-80
:MEASure:SQFormula
Function
Sets the equation used to compute S (apparent
power) and Q (reactive power) or queries the
current setting.
:MEASure:SQFormula {TYPE1|TYPE2|
Syntax
TYPE3}
:MEASure:SQFormula?
:MEASURE:SQFORMULA TYPE1
Example
:MEASURE:SQFORMULA? ->
:MEASURE:SQFORMULA TYPE1
Description • For details on the equation corresponding
to {TYPE1|TYPE2|TYPE3} , see the User's
Manual IM WT3001E-01EN.
• "TYPE3" is selectable only on models with the
advanced computation function (/G6 option).
:MEASure:SYNChronize
Function
Sets the synchronized measurement mode or
queries the current setting.
:MEASure:SYNChronize {MASTer|SLAVe}
Syntax
:MEASure:SYNChronize?
:MEASURE:SYNCHRONIZE MASTER
Example
:MEASURE:SYNCHRONIZE? ->
:MEASURE:SYNCHRONIZE MASTER
IM WT3001E-17EN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt3002eWt3003eWt3004e

Table of Contents