6.7 DISPlay Group
:DISPlay:WAVE:MAPPing:{U<x>|I<x>|SPE
ed|TORQue|MATH<x>}
Function
Sets the mapping of the {voltage|current|rotating
speed|torque|waveform computation} waveform
to the split screen or queries the current setting.
:DISPlay:WAVE:MAPPing:{U<x>|I<x>|
Syntax
SPEed|TORQue|MATH<x>} {<NRf>}
:DISPlay:WAVE:MAPPing:{U<x>|I<x>|
SPEed|TORQue|MATH<x>}?
<x> of U<x>, I<x> = 1 to 4 (element)
<x> of MATH<x> = 1 to 2 (MATH)
<NRf> = 0 to 3
:DISPLAY:WAVE:MAPPING:U1 0
Example
:DISPLAY:WAVE:MAPPING:U1? ->
:DISPLAY:WAVE:MAPPING:U1 0
Description • This command is valid when the waveform
mapping method ( :DISPlay:WAVE:
MAPPing[:MODE] ) is set to "USER."
• {SPEed|TORQue} are valid only on models
with the motor evaluation function (/MTR
option).
• MATH<x> is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:WAVE:POSition?
Function
Queries all settings related to the vertical position
(level of the center position) of the waveform.
:DISPlay:WAVE:POSition?
Syntax
:DISPLAY:WAVE:POSITION? ->
Example
:DISPLAY:WAVE:POSITION:U1 0.000;
U2 0.000;U3 0.000;U4 0.000;
I1 0.000;I2 0.000;I3 0.000;I4 0.000
:DISPlay:WAVE:POSition:{UALL|IALL}
Function
Collectively sets the vertical position (level of the
center position) of the waveform {voltage|current}
of all elements.
:DISPlay:WAVE:POSition:{UALL|
Syntax
IALL} {<NRf>}
<NRf> = –130.000 to 130.000(%)
:DISPLAY:WAVE:POSITION:UALL 0
Example
:DISPlay:WAVE:POSition:{U<x>|I<x>}
Function
Sets the vertical position (level of the center
position) of the waveform {voltage|current} of the
element or queries the current setting.
:DISPlay:WAVE:POSition:{U<x>|
Syntax
I<x>} {<NRf>}
:DISPlay:WAVE:POSition:{U<x>|I<x>}?
<x> = 1 to 4 (element)
<NRf> = –130.000 to 130.000(%)
:DISPLAY:WAVE:POSITION:U1 0
Example
:DISPLAY:WAVE:POSITION:U1? ->
:DISPLAY:WAVE:POSITION:U1 0.000
6-42
:DISPlay:WAVE:SVALue (Scale VALue)
Function
Turns ON/OFF the scale value display or queries
the current setting.
:DISPlay:WAVE:SVALue {<Boolean>}
Syntax
:DISPlay:WAVE:SVALue?
:DISPLAY:WAVE:SVALUE ON
Example
:DISPLAY:WAVE:SVALUE? ->
:DISPLAY:WAVE:SVALUE 1
:DISPlay:WAVE:TDIV
Function
Sets the Time/div value of the waveform or
queries the current setting.
:DISPlay:WAVE:TDIV {<Time>}
Syntax
:DISPlay:WAVE:TDIV?
<Time> = 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500
(ms), 1, or 2 (s)
:DISPLAY:WAVE:TDIV 5MS
Example
:DISPLAY:WAVE:TDIV? ->
:DISPLAY:WAVE:TDIV 5.0E-03
Description The specifiable Time/div value is up to 1/10 of the
data update interval ( :RATE ).
:DISPlay:WAVE:TLABel (Trace LABel)
Function
Turns ON/OFF the waveform label display or
queries the current setting.
:DISPlay:WAVE:TLABel {<Boolean>}
Syntax
:DISPlay:WAVE:TLABel?
:DISPLAY:WAVE:TLABEL OFF
Example
:DISPLAY:WAVE:TLABEL? ->
:DISPLAY:WAVE:TLABEL 0
:DISPlay:WAVE:TRIGger?
Function
Queries all settings related to the trigger.
:DISPlay:WAVE:TRIGger?
Syntax
:DISPLAY:WAVE:TRIGGER? ->
Example
:DISPLAY:WAVE:TRIGGER:MODE AUTO;
SOURCE U1;SLOPE RISE;LEVEL 0.0
:DISPlay:WAVE:TRIGger:LEVel
Function
Sets the trigger level or queries the current
setting.
:DISPlay:WAVE:TRIGger:LEVel {<NRf>}
Syntax
:DISPlay:WAVE:TRIGger:LEVel?
<NRf> = –100.0 to 100.0 (%) (The resolution is
0.1(%))
:DISPLAY:WAVE:TRIGGER:LEVEL 0
Example
:DISPLAY:WAVE:TRIGGER:LEVEL? ->
:DISPLAY:WAVE:TRIGGER:LEVEL 0.0
Description Set the value in terms of a percentage of the full
scale value displayed on the screen.
IM WT3001E-17EN