YOKOGAWA WT3001E User Manual page 150

Precision power analyzer communication interface
Hide thumbs Also See for WT3001E:
Table of Contents

Advertisement

6.21 STORe Group
:STORe:INTerval
Function
Sets the store interval or queries the current
setting.
:STORe:INTerval {<NRf>,<NRf>,<NRf>}
Syntax
:STORe:INTerval?
1st <NRf> = 0 to 99 (hour)
2nd <NRf> = 0 to 59 (minute)
3rd <NRf> = 1 to 59 (second)
:STORE:INTERVAL 0,0,0
Example
:STORE:INTERVAL? ->
:STORE:INTERVAL 0,0,0
:STORe:ITEM
Function
Sets the stored item or queries the current
setting.
:STORe:ITEM {NUMeric|WAVE|NWAVe}
Syntax
:STORe:ITEM?
NUMeric = Store only the numeric data.
WAVE = Store only the waveform display data
NWAVe = Store both the numeric data and
waveform display data
:STORE:ITEM NUMERIC
Example
:STORE:ITEM? -> :STORE:ITEM NUMERIC
:STORe:MEMory?
Function
Queries all settings related to the storage
memory.
:STORe:MEMory?
Syntax
:STORE:MEMORY? ->
Example
:STORE:MEMORY:ALERT 1
:STORe:MEMory:ALERt
Function
Sets whether to display a confirmation message
when clearing the storage memory or queries the
current setting.
:STORe:MEMory:ALERt {<Boolean>}
Syntax
:STORe:MEMory:ALERt?
:STORE:MEMORY:ALERT ON
Example
:STORE:MEMORY:ALERT? ->
:STORE:MEMORY:ALERT 1
Description The initialization takes place immediately
when initializing the storage memory using the
" :STORe:MEMory:INITialize " command
regardless of the setting specified with this
command.
:STORe:MEMory:CONVert:ABORt
Function
Abort converting the stored data from the memory
to the file.
:STORe:MEMory:CONVert:ABORt
Syntax
:STORE:MEMORY:CONVERT:ABORT
Example
6-106
:STORe:MEMory:CONVert:EXECute
Function
Executes the converting of the stored data from
the memory to the file.
:STORe:MEMory:CONVert:EXECute
Syntax
:STORE:MEMORY:CONVERT:EXECUTE
Example
Description • The convert destination file is set using the
" :STORe:FILE:... " command.
• When file conversion is executed, this
instrument accesses the file twice.
To confirm the completion of the file conversion,
use the " COMMUNICATE:WAIT 64 " command
(checks the change in bit 6 (ACS) of the
condition register) and check the completion
of the file access of this instrument twice. An
example is indicated below.
"STATUS:EESR?"
(Clear the extended event register)
"STORE:MEMORY:CONVERT:EXECUTE"
(Start the file conversion)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the first time)
"STATUS:EESR?"
(Clear the extended event register)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the second
time)
"STATUS:EESR?"
(Clear the extended event register)
:STORe:MEMory:INITialize
Function
Executes the initialization of the storage memory.
:STORe:MEMory:INITialize
Syntax
:STORE:MEMORY:INITIALIZE
Example
:STORe:MODE
Function
Sets the data storage/recall or queries the current
setting.
:STORe:MODE {STORe|RECall}
Syntax
:STORe:MODE?
:STORE:MODE STORE
Example
:STORE:MODE? -> :STORE:MODE STORE
IM WT3001E-17EN