6.7 DISPlay Group
:DISPlay:TRENd:ITEM<x>:SCALing:MODE
Function
Sets the scaling mode of the trend or queries the
current setting.
:DISPlay:TRENd:ITEM<x>:SCALing:
Syntax
MODE {AUTO|MANual}
:DISPlay:TRENd:ITEM<x>:SCALing:
MODE?
<x> = 1 to 16 (item number)
:DISPLAY:TREND:ITEM1:SCALING:
Example
MODE AUTO
:DISPLAY:TREND:ITEM1:SCALING:MODE?
-> :DISPLAY:TREND:ITEM1:SCALING:
MODE AUTO
:DISPlay:TRENd:ITEM<x>:SCALing:VALue
Function
Sets the upper and lower limits of manual scaling
of the trend or queries the current setting.
:DISPlay:TRENd:ITEM<x>:SCALing:
Syntax
VALue {<NRf>,<NRf>}
:DISPlay:TRENd:ITEM<x>:SCALing:
VALue?
<x> = 1 to 16 (item number)
<NRf> = –9.999E+30 to 9.999E+30
:DISPLAY:TREND:ITEM1:SCALING:
Example
VALUE 100,-100
:DISPLAY:TREND:ITEM1:SCALING:VALUE?
-> :DISPLAY:TREND:ITEM1:SCALING:
VALUE 100.0E+00,-100.0E+00
Description • Set the upper limit and then the lower limit.
• This command is valid when the scaling mode
of the trend
(:DISPlay:TRENd:ITEM<x>:SCALing:
MODE) is set to "MANual."
:DISPlay:TRENd:TDIV
Function
Sets the horizontal axis (T/div) of the trend or
queries the current setting.
:DISPlay:TRENd:TDIV {<NRf>,<NRf>,
Syntax
<NRf>}
:DISPlay:TRENd:TDIV?
{<NRf>, <NRf>, <NRf>} = 0, 0, 3 to 24, 0, 0
1st <NRf> = 1, 3, 6, 12, 24 (hour)
2nd <NRf> = 1, 3, 6, 10, 30 (minute)
3rd <NRf> = 3, 6, 10, or 30 (second)
:DISPLAY:TREND:TDIV 0,0,3
Example
:DISPLAY:TREND:TDIV? ->
:DISPLAY:TREND:TDIV 0,0,3
Description Set the three <NRf>'s so that one <NRf> is a
non-zero value and the other two are zeroes.
6-40
:DISPlay:TRENd:T<x>
Function
Turns ON/OFF the trend or queries the current
setting.
:DISPlay:TRENd:T<x> {<Boolean>}
Syntax
:DISPlay:TRENd:T<x>?
<x> = 1 to 16 (item number)
:DISPLAY:TREND:T1 ON
Example
:DISPLAY:TREND:T1? ->
:DISPLAY:TREND:T1 1
:DISPlay:VECTor?
Function
Queries all settings related to the vector display.
:DISPlay:VECTor?
Syntax
:DISPLAY:VECTOR? -> :DISPLAY:
Example
VECTOR:OBJECT SIGMA;NUMERIC 1;
UMAG 1.000;IMAG 1.000
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:VECTor:NUMeric
Function
Turns ON/OFF the numeric data display for the
vector display or queries the current setting.
:DISPlay:VECTor:NUMeric {<Boolean>}
Syntax
:DISPlay:VECTor:NUMeric?
:DISPLAY:VECTOR:NUMERIC ON
Example
:DISPLAY:VECTOR:NUMERIC? ->
:DISPLAY:VECTOR:NUMERIC 1
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:VECTor:OBJect
Function
Sets the wiring unit to be displayed during vector
display or queries the current setting.
:DISPlay:VECTor:OBJect {SIGMA|
Syntax
SIGMB}
:DISPlay:VECTor:OBJect?
:DISPLAY:VECTOR:OBJECT SIGMA
Example
:DISPLAY:VECTOR:OBJECT? ->
:DISPLAY:VECTOR:OBJECT SIGMA
Description • This command is valid only on models with the
advanced computation function (/G6 option).