4.3 Setting the Ethernet Control
Explanation
4-6
You can control this instrument from a PC using the Ethernet interface. YOKOGAWA's
dedicated library software (TMCTL) must be installed on the PC in addition to entering
the settings described above.
Retail Software
WTViewer (760122)
You can use WTViewer by setting this instrument as WT3000 emulate mode.
A trial version is available for download from the following URL.
http://tmi.yokogawa.com/products/digital-power-analyzers/power-measurement-application-
software/wtviewer-760122/
Setting the User Name
• Enter the user name to allow access to this instrument.
• Enter up to 15 characters.
• The characters that can be used are 0-9, A-Z, %, _, ( ) (parentheses), - (minus sign).
• If you specify anonymous, this instrument can be accessed from the outside (PC)
without a password.
Setting the Password
• Enter the password for the user name to allow access to this instrument.
• Enter up to 15 characters.
• The characters that can be used are 0-9, A-Z, %, _, ( ) (parentheses), - (minus sign).
• If the user name is set to anonymous, this instrument can be accessed from the
outside (PC) without a password.
• The password setting is entered twice.
Setting the Timeout Value
This instrument closes the connection to the network if there is no access for a certain
period of time (timeout time).
The available settings are 0 to 3600 s, or Infinite. The default value is Infinite.
Note
To apply new settings, this instrument must be power cycled.
IM WT3001E-17EN