5.15 INTEGrate Group

The commands in this group deal with integration.
You can make the same settings and queries that you can on the screen that appears by using Setup menu >
Computation/Output tab > Integration button.
:INTEGrate?
Function
Queries all integration settings.
:INTEGrate?
Syntax
:INTEGrate:ACAL
Function
Sets or queries the on/off status of integration
auto calibration.
:INTEGrate:ACAL {<Boolean>}
Syntax
:INTEGrate:ACAL?
:INTEGRATE:ACAL OFF
Example
:INTEGRATE:ACAL?
-> :INTEGRATE:ACAL 0
:INTEGrate:INDependent
Function
Sets or queries the on/off status of independent
element integration.
:INTEGrate:INDependent {<Boolean>}
Syntax
:INTEGrate:INDependent?
:INTEGRATE:INDEPENDENT OFF
Example
:INTEGRATE:INDEPENDENT?
-> :INTEGRATE:INDEPENDENT 0
:INTEGrate:MODE
Function
Sets or queries the integration mode.
:INTEGrate:MODE {NORMal|CONTinuous|
Syntax
RNORmal|RCONtinuous}
:INTEGrate:MODE?
NORMal = Normal integration mode
CONTinuous = Continuous integration mode
RNORmal = Real-time normal integration mode
RCONtinuous = Real-time continuous integration
mode
:INTEGRATE:MODE NORMAL
Example
:INTEGRATE:MODE?
-> :INTEGRATE:MODE NORMAL
:INTEGrate:QMODe?
Function
Queries the current modes for current integration
of all elements.
:INTEGrate:QMODe?
Syntax
:INTEGrate:QMODe[:ALL]
Function
Collectively sets the current mode for current
integration of all elements.
:INTEGrate:QMODe[:ALL] {RMS|MEAN|DC|
Syntax
RMEAN|AC}
:INTEGRATE:QMODE:ALL DC
Example
5-92
:INTEGrate:QMODe:ELEMent<x>
Function
Sets or queries the current mode for current
integration of the specified element.
:INTEGrate:QMODe:ELEMent<x> {RMS|
Syntax
MEAN|DC|RMEAN|AC}
:INTEGrate:QMODe:ELEMent<x>?
<x> = 1 to 7 (element)
:INTEGRATE:QMODE:ELEMENT1 DC
Example
:INTEGRATE:QMODE:ELEMENT1?
-> :INTEGRATE:QMODE:ELEMENT1 DC
Description Regardless of the independent element
integration setting (:INTEGrate:INDependent),
this instrument operates according to the current
mode of the specified element.
:INTEGrate:RACTion
Function
Sets or queries the integration resume action that
is taken when the power recovers from a power
failure while integration is in progress.
:INTEGrate:RACTion {STARt|STOP|ERRor}
Syntax
:INTEGrate:RACTion?
STARt = Integration is resumed automatically.
STOP = Integration is placed in a stopped state.
Integration can be resumed manually (START
key or command).
ERRor = Integration error state occurs. Resuming
of integration is prohibited.
:INTEGRATE:RATION STOP
Example
:INTEGRATE:RACTION?
-> :INTEGRATE:RACTION STOP
:INTEGrate:RESet
Function
Resets the integrated value.
:INTEGrate:RESet {[<NRf>][,<NRf>]
Syntax
[,<NRf>][,<NRf>][,<NRf>][,<NRf>]
[,<NRf>]}
<NRf> = 1 to 7 (element)
:INTEGRATE:RESET (resets all elements)
Example
:INTEGRATE:RESET 1,2,3 (resets the
specified elements)
Description • When independent element integration
(:INTEGrate:INDependent) is set to ON (1),
you can use the parameters to specify which
elements to perform the operation on. If you
omit the parameters, the operation will be
performed on all the elements.
• When independent element integration
(:INTEGrate:INDependent) is set to OFF (0),
you cannot use the parameters to specify
elements.